Syringe pump KSP–A200 is a single-channel light–weight, infusion system designed for low volume infusion. Programmed infusion modes, administer drug volume at specified rate as per patient dosage. Dose titration allows uninterrupted flow during change of flow rate. Real-time pressure sensing with anti-bolus and purge feature assures safe operation.
Ask Quote
Channel | Single channel infusion |
---|---|
Volume (VTBI) | 0 ml – 9999 ml |
Injection Compliance | Adaptable for 6 syringe sizes |
Accuracy | ≤ ±2 % |
KVO Rate | 0.01 ml/h – 5 ml/h (steps : 0.01 ml/h) |
Occlusion System | 12 selectable occlusion levels |
Syringe Size | 5 ml, 10 ml, 20 ml, 30 ml, 50 ml, 60 ml |
Infusion Mode | Rate mode/Time mode, Body weight, TIVA, Relay mode Loading-dose, Ramp up/down, Sequence, Drug library mode, Micro mode |
Flow Selection | 0.01 ml/h ~ 1500 ml/h, increment 0.01 ml/h |
Bolus Rate | 0.1 ml/h ~ 1500 ml/h (Auto or Manual bolus control) |
Display | 4.3-inch colour LCD screen |
Battery | Lithium polymer battery 11.1 V 2600 mAh About 12 hours of operation at 5ml/h (< 5 hours of charging) |
Memory | Records over 5000 entries for review |
Drug library | Database of up to 2000 drugs |
Safety | Class I Type CF |
Moisture Protection | IP24 |
Operation | Ambient Temperature: 5 °C to 40 °C Relative Humidity: 20 % to 90 % |
Storage | Ambient Temperature: –30 °C to 55 °C Relative Humidity: ≤ 95% |
Power Supply | AC : 110 V ~ 240 V, 50/60 Hz DC : 12 V |
Dimensions | 235 × 90 × 120 mm |
Weight | 1.7 kgs |